Product Overview:

The ROCHE Cobas 6000 Chemistry Analyzer is the first member of the new cobas modular platform. It offers medium workload laboratories tailor-made solutions for clinical chemistry and immunochemistry testing. The cobas® 6000 analyzer series can consolidate into one integrated platform up to 95% of your routine workload of clinical chemistry and immunoassay testing. The cobas® 6000 analyzer series automates and simplifies your lab’s workload. Fully compatible with MODULAR PRE-ANALYTICS PLUS system, the cobas® 6000 analyzer series can seamlessly integrate pre-analytical automation. We’ve made operations in your lab simpler by standardizing Roche Diagnostics analyzers with common reagent carriers, applications, analytical technologies, and software interface. With the cobas® 6000 analyzer series, there is one sample tube, one patient result report, and one LIS connection. The cobas® 6000 analyzer series gives you the power to control your lab, improve financial performance, enhance efficiency, and make a difference for the doctors and patients who rely on the critical information you provide to make vital decisions about healthcare, every day.

System:

2nd Generation SWA: modular, analytical system platform, consolidated work area for Clinical Chemistry and Immunology, expandable and re-configurable on site

Types of Modules:

  • cobas c 501 module: Photometric measuring unit (incl. ISE)
  • cobas e 601 module: Electrochemiluminescence (ECL) technology measuring unit

Number of module combinations:
7 module combinations, Up to 3 modules in one core unit

  • cobas c 501 modules 1 to 2 modules
  • cobas e 601 modules 1 to 2 modules
  • cobas c 501 | cobas e 601 modules - 2 to 3 modules (only 2 unites of 1 module type for SWA combinations possbile)

Sample throughput:

Up to 120 racks/hr or 600 samples/hr

Test throughput (theoretical max):

  • 170 up to 2,170 tests/hr - overall
  • 1,000 up to 2,000 tests/hr - with cobas c 501 modules
  • 170 up to 340 tests/hr - with cobas e 601 modules
  • 1,170 up to 2,170 tests/hr - with cobas c 501 and cobas e 601 modules

Number of channels (reagent slots):

  • Up to 151 channels, in total
  • 63 channels (including ISE, automatic setting) on cobas c 501 module
  • 25 channels on cobas e 601 module

Sample material:

Serum, Plasma, Urine, CSF

Core unit analytics:

  • Rack: 5 position rack, RD standard rack
  • Rack types: Routine, STAT, Control, Calibrator, Rerun (manual), Wash
  • Tray: Tray with 15 racks/75 samples, RD standard tray
  • STAT port: STAT samples are processed with true STAT interrupt

Sample container types:

  • Primary tubes:5 to 10 mL; 16 x 100, 16 x 75, 13 x 100, 13 x 75 mm
  • Sample cup: 2.5 mL
  • Micro cup: 1.5 mL, (exception: cobas e 601 module)
  • Cups on tube: Cup on top of a 16 x 75/100 mm tube; Cup on top of one non standard tube
  • False bottom tube (FBT): One type is definable

Sample volume:

1 - 35 µL

Sample dilution:

3 - 121 times

Sample clot detection:

Available for cobas c 501 and cobas e 601 module

Minimum sample volume:

  • Primary tubes: 700 µL
  • Sample cup: 100 µL
  • Micro cup (FTB): 50 µL

Sample barcode types:

Code 128; Codabar (NW 7); Interleaved 2 of 5; Code 39

System interfaces:

  • RS 232 serial interface, bi-directional
  • Interface to cobas link data station for cobas teleservice functionalities and automatic download

Sample data base:

10.000 routine/STAT samples

Test methods:

For photometric modules: 1 point, 1 point + prozone check, 2 point, 2 point kinetic, 2 point + prozone check, 3 point, 1 point + kinetics Rate A, Rate A + serum index, Rate A with blank, Rate B

Calibrator/QC input:
Via specified racks through the input buffer or STAT port

Calibration methods:

  • Start-up, Re-calibration
  • For photometric modules: Linear, non-linear multi-points, 2 point calibration, K-factor up to 100 different calibrators pre-programmable
  • Storage of up to 180 curves
  • Preventive calibration of the stand-by cobas c and cobas e packs on cobas c 501: Two k-factor can be defined for different sample types

QC methods:

  • For e- and photometric modules: Real-time QC, individual QC, cumulative QC
  • Up to 100 controls pre-programmable
  • Preventive QC after calibration of the stand-by cassettes/rack packs
  • Auto QC: QC without operator intervention (timer triggered)

Rerun/reflex function:

  • Automatic rerun and manual rerun
  • Automatic reflex is supported by the system, reflex request to be provided by Roche Middleware Solutions

Electrical requirements:

  • Power requirements: 230 Volts AC; 110 Volts AC 2.4 kVA (for cobas c 501 | e 601)
  • Frequency: 50 Hz or 60 Hz +/- 0.5 %

Water/waste requirements:

  • Water: Bacteria free, deionised water supply: resistance of < 1,0 µS/cm
  • Water pressure: 0.5 – 3.5 kg/cm3, (49 - 343 Kpa)
  • Biohazards waste: Separate container behind the system
  • L. concentrated: Central drain port, diameter: = 50 mm / 2 inches
  • Waste drain < waste: 100 mm above the floor / 4 inches

Water consumption:

  • Max flow rate: cobas c 501 module: approx. 40 L/hr
  • Typical consumption: cobas e 601 module: approx. 30 L/hr

Operating conditions:

  • Ambient temperature: 15 to 32 °C / 59 to 90 °F
  • Ambient humidity: 30 to 85 % (RH, without condensation)
  • Heat output: 1.5 kW (5400 kJ/hr for cobas c 501, 4320 kJ/hr for cobas e 601)
  • Noise output: < 65 dB